R&S SMB Module Replacement
1407.0806.82 3.39 E-6
Dismounting the Case
Put the instrument on the front shock mounts.
Unscrew the four screws of the rear shock mounts (rear right and rear left) on both sides and take
them off.
Pull off the case.
When mounting the case, take care not to damage or pull off cables!
Replacing Fuses
NOTICE Risk of damage to the instrument
Ensure that the power supply cord is disconnected.
If you are not using the supplied spares, be sure to use fuses of the same type and
rating.
Switch off the instrument.
Disconnect the power supply cord.
To replace a fuse, unscrew the fuse holder at the rear panel of the instrument.