Test Procedures R&S SMB
1407.0806.82 1.48 E-5
Stereo/RDS Coder (Option R&S SMB-B5)
Frequency Response
Test setup DUT RF output connected to Modulation Analyzer RF input, Audio
analyzer LF output connected to DUT Stereo L, R input.
Test method
Measurement of demodulated AF signal by Modulation Analyzer
Measurement
Setting on DUT:
RF On
Frequency 100 MHz
Level 0 dBm
Stereo Modulation menu:
State On
FM Deviation 40 kHz
Audio Source LF Gen (EXT L, R)
Mode Mono L (Mono R)
LF Gen Frequency 20 Hz
LF Gen Shape Sine
Stereo Pilot tone settings:
Pilot State On
Pilot Deviation 6.75 kHz
ARI settings:
State Off
RDS settings:
State Off
Setting on Audio Analyzer UPL:
GEN:
INSTRUMENT ANALOG: Channel(s) 1 (2)
FUNCTION Sine
FUNCTION Frequency = 20 Hz
FUNCTION Voltage = 0.707 V
Setting on Modulation Analyzer FMB:
DEMODULATION FM-STEREO
CHANNEL L (R)
RELATIVE
Vary the LF Generator frequency and measure the reference level
(MEAS-REF).
Recommended settings
LF in logarithmic steps, 7 steps per decade (1, 1.5, 2, 3, 4, 6, 8)
from 20 Hz to 15 kHz.
Repeat the measurement for all four combinations (internal L,
internal R, external L, external R).