R&S SMB Module Replacement
1407.0806.82 3.47 E-6
Replacing the Basis Board (A100)
(See chapter 5, drawing 1406.6000.01).
Note: Different Basis Boards, order no. 1406.6600.02 or 1406.6700.02.
Removing the Basis Board
Switch off the instrument and pull the mains plug.
Dismounting the case (see page 3.39).
Unfasten lock of fan cable
Disconnect the fan cable from X116 and the power supply cable from X101 (Remove the plastic tie
before disconnect the cable).
Disconnect the flex. cables W330, W310, W222 and the flex. switchboard cable.
Unscrew one combination screw (0041.1653.00) and two screws (0041.1653.00) from the
IEC-connector at the back of the instrument.
Undo the five combination screws (red marked) on the basis board.
Remove the basis board (by turning it up towards the front and pulling it out and upward).
Remove the SIM Card from the basis board (see page 3.50)