R&S SMB Test Procedures
1407.0806.82 1.47 E-5
Pulse Generator
PULSE VIDEO
Test equipment - Storage oscilloscope (Table 1-1, item 3) with 50V Feed-through
termination on input.
Test setup
Connect the PULSE VIDEO socket on the rear of the DUT to the
storage oscilloscope.
Measurement
Setting on DUT:
Pulse Modulation menu:
State On
Source: Pulse Generator
Pulse Generator menu:
State On
Pulse Period 10 Zs
Pulse Width 5 Z s
Setting on oscilloscope:
1 V/div
Time base 2.5 Zs/div
Trigger: 50 % of amplitude, rising edge.
Check the signal for a symmetric square wave with 10 Zs pulse
period and 3 V amplitude.
Rise and fall time < 10 ns