Test Procedures R&S SMB
1407.0806.82 1.46 E-5
Video Crosstalk
Test assembly As above for ON/OFF Ratio
Measurement
Setting on pulse generator:
Square wave pulse sequence with a frequency of 100 kHz,
TTL level
Setting on DUT:
RF On
Frequency 1 GHz, 6 GHz
Level 0 dBm
Pulse Modulation State On
Settings on the Analyzer
REF LEVEL 0 dBm
FREQ CENTER 100 kHz
SPAN 10 kHz
Measure the signal level at 100 kHz with the analyzer.
The Video Crosstalk is the amplitude of the spectral line found
at 100 kHz related to the RF carrier level