Test Procedures R&S SMB
1407.0806.82 1.24 E-5
Example:
R
ef 0 %
1 AP
C
LRWR
A
I
FB 2 MHz
Center 1 GHz 20 ms/
A
QT 200 ms
A
tt 5 dB
*
-80
-60
-40
-20
0
20
4
0
60
8
0
Amplitude Modulation Summary
Modulation Depth 0.022 %
Modulation +peak 0.022 %
-peak -0.023 %
±
peak/2 0.022 %
RMS 0.005 %
Carrier Power -0.87 dBm
Modulation Frequency --- Hz
Sampling Rate 250 kHz
Record Length 50001
Demod Bandwidth 200 kHz
AF Filter HP 20 Hz
LP 23 kHz
Date: 25.FEB.2008 16:33:44
Residual AM = 0.005 %