Test Procedures R&S SMB
1407.0806.82 1.30 E-5
Recommended test levels for
Level Accuracy
3 mV, 10 mV, 30 mV, 100 mV, 300 mV, 1 V and 3 V
Measurement of Frequency
response
Settings on DUT:
LF Output menu:
LF Output Voltage 1 V
set LF Gen Frequency to recommended test frequencies and
measure the output level
Determine the highest and the lowest level V
max
and V
min
.
The frequency response in dB is defined as:
D = 20*log
10
(V
max
) - 20*log
10
(V
min
)
Recommended test
frequencies
10 Hz, 1 kHz, 10 kHz, 100 kHz and 1 MHz
Note: The settling time is a pure computer time and needs therefore not to be measured.