EasyManua.ls Logo

Rohde & Schwarz SMB100A - Page 65

Rohde & Schwarz SMB100A
223 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test Procedures R&S SMB
1407.0806.82 1.42 E-5
PhiM Distortion
Test Method
FFT Demodulation (see chapter "Test ")
Measurement
Settings on DUT:
RF On
Level 0 dBm
Phase Modulation menu:
State on
PhiM Source Internal
PhiM Mode Normal
LFGen Frequency 10 kHz
Settings on R&S FSQ:
see chapter "Test "
SWEEP / MEAS TIME 2.5 ms,
FREQ / AF STOP 50 kHz,
AMPTD / REF LEVEL 0 dBm,
FREQ / CENTER test frequency.
Read THD from the Display. To convert to percent calculate
THDpct = 100 * 10 ^ (THDdB/20).
Recommended settings
CF sweep
Recommended test frequencies
with PhiM deviation 1 rad: 10 MHz, 375.1 MHz, 500 MHz, 625 MHz,
750 MHz,
with PhiM deviation 2 rad: 1100 MHz,
with PhiM deviation 4 rad: 2200 MHz,
with PhiM deviation 8 rad: 3200 MHz, 6 GHz
Repeat measurement with changing settings to
PhiM Mode Low Noise
PhiM deviation 2.5 rad
for test frequencies: 375.1 MHz, 750 MHz
Repeat measurement with changing settings to
PhiM Mode High Deviation
PhiM deviation 10 rad
for test frequencies: 375.1 MHz, 750 MHz

Table of Contents

Other manuals for Rohde & Schwarz SMB100A

Related product manuals