Test Procedures R&S SMB
1407.0806.82 1.50 E-5
Measurement
Setting on Modulation Analyzer FMB:
DEMODULATION FM-STEREO
CHANNEL L (R)
NOISE FILTER OFF
ABSOLUTE
MAN RANGE dBm
RANGE 12 dBm
AUTO DIST-SINAD
Channel separation: First note the level of the AF signal on the UPL
in channel 1 as reference, then switch the source to channel 2 and
measure level of crosstalk 2 -> 1. Carry out the same in reverse
and measure level of crosstalk 1 -> 2.
Harmonic distortion: Read the harmonic distortion value from
Modulation Analyzer FMB.
Recommended settings
CF sweep
Recommended test frequencies: 66 MHz, 76 MHz, 87 MHz, 98
MHz, 110 MHz