80
The value entered for the gravitational acceleration at the place of installation is
displayed in m/s
2
.
In the example shown here, the gravitational acceleration for the setting
“Germany (Zone D)” is displayed.
Confirm the value displayed for gravitational acceleration, or cancel the
) calibration procedure. Press ( to cancel.
You are prompted to place the required weight on the platform (e.g., 5.0 kg).
The subsequent steps for completing the calibration/adjustment are described in
the chapter entitled “Operation,” under “Calibration and Adjustment.”
At the conclusion of A/D converter configuration, return the menu access switch from
the “accessible” to the “blocked” position:
§ If necessary, remove the cap that covers the menu access switch on the left-hand side
of the back of the indicator.
§ Move the menu access switch to the left (“blocked” position).
See also “Calibration and Adjustment” in the chapter entitled “Operation,”
or refer to the service manual.
§ Replace the protective cap over the menu access switch.
ee Restart the weighing instrument: Turn the indicator off and then on again.
The Sartorius logo is displayed briefly, after which the device is in normal weighing mode.
If the A/D converter was configured with a “Verifiable” data record, the lines for display
of metrological data (lines 1 and 2) show the data valid for use in legal metrology,
if the menu access switch is closed. For details, see the section entitled “Checking
and Configuring the Equipment for Use in Legal Metrology” in this chapter.