EasyManuals Logo

Sartorius CW3P User Manual

Sartorius CW3P
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
83
Soft key O
If nec.: soft key Q | q, soft key O
The Setup menu for the “WP 1 - Internal” device parameters is displayed.
Soft key q, soft key O Open the Calibration/adjustment menu.
The “Calibration/adjustment” submenu for the selected weighing platform (in this
example, “WP 1 - Internal”) is displayed.
Soft key O Open the CAL key function submenu.
The “CAL Key Function” submenu is displayed.
Important Note:
Which functions can be configured in the “CAL Key Function” submenu depends on the
selected weighing platform and its configuration data. Functions that cannot be activated
are not displayed in the selection list (in this example, the “Ext. lineariz.; factory-def. wt.”
function).
Important Note:
When you press the CAL key to perform the calibration/adjustment function you have
defined here, the function will be carried out in the normal weighing mode (the display
shows
CAL), because the Service mode is deactivated when you exit the Setup menu.
To perform the function on a digital weighing platform (such as an IS platform),
however, it must be carried out in Service mode. The procedure for this is as follows:
after selecting the desired function in the “CAL Key Function” menu and exiting the
Setup menu, reactivate the Service mode again and then exit the Setup menu immediately
by pressing M or the oo soft key.
The weighing instrument is now in Service mode, even though this is not indicated
on the display. Press and hold the ) key at least 2 seconds to activate the selected
calibration/adjustment function. The display shows
S-CAL, indicating that the scale
is in Service mode. If you cancel the function by pressing (, or restart the scale by
pressing e, the Service mode is deactivated.
External Calibration/Adjustment with Factory-Defined Weight (Default Weight)
If nec.: Soft key Q, soft key l If not already selected (factory setting, marked by o when active), select menu item
Ext. cal./adjust.; factory-defined cal weight
(external calibration/adjustment with the weight value configured at the factory).
To do this, press the Q soft key (repeatedly if necessary) to highlight this menu item and
press the l soft key to confirm. The menu item is marked by a circle (o).
Important Note:
The menu items “Ext. cal./adjust.; factory-defined cal wt.” (external calibration/adjust-
ment using the weight value defined at the factory – default weight), “Ext. cal./adjust.;
user-defined weight” (external calibration/adjustment with a user-defined weight) and
“Cal key blocked” are also accessible without activating the Service mode.
ee Restart the scale: Turn the indicator off and then on again.
The Sartorius logo is displayed briefly, after which the device is in normal weighing mode.