Page #84 background imageLoading...
Page #84 background image
84
If nec.: ( Unload and zero the scale.
) (> 2 sec) Start external calibration/adjustment.
This display is shown for 2 seconds.
Important Note:
If the display of geographical data (elevation and latitude or gravitational acceleration) is
activated (see “Settings” at the beginning of this section), this data is displayed; press )
to confirm each value (to cancel the calibration/adjustment procedure, press ().
For details, see “Entering Geographical Data” above.
The nominal value of the required calibration weight (in this example, 5.000 kg) is shown
as a negative value on the display.
§ Place the required weight on the scale.
Important Note:
If the calibration/adjustment sequence is set to “Cal then auto adjust,” (menu path
“Calibration/adjustment > Cal./adj. sequence > Cal. then auto adj.”; see “Settings” at the
beginning of this section) and the calibration weight consists of more than one weight,
apply the weights to the scale in series at short intervals. When the weighing instrument
has stabilized, the weight on the scale is accepted as the calibration weight after
a predefined interval, and the weighing instrument is calibrated/adjusted with this
weight. The difference since the most recent span adjustment is not displayed; this value
is output only on GMP-compliant printouts (see next page).
After a brief pause, the difference since the last span adjustment is displayed (calibration).
Important Note:
This value is displayed only if the setting “Cal. then manual adjust” is active (see previous
“Note”). If “Cal. then auto adj.” is active, the calibration/adjustment procedure
cannot be cancelled.
$ To stop the procedure after calibration and before adjustment takes place, press (
(only if “Cal. then manual adj.” is active).
) Perform adjustment (only if “Cal. then manual adj.” is active).
At the conclusion of the calibration procedure, the calibration weight is displayed as a
positive value.
——
CAL

Other manuals for Sartorius CW3P

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius CW3P and is the answer not in the manual?

Sartorius CW3P Specifications

General IconGeneral
BrandSartorius
ModelCW3P
CategoryAccessories
LanguageEnglish

Related product manuals