91
Display after the scale has been zeroed.
§ Place the preload weight on the weighing platform.
) (> 2 sec) Activate the “Set preload” function.
This display is shown for 2 seconds. The weight on the scale is stored as the preload.
After the “Set Preload” operation has been completed, the scale is zeroed.
The scale then returns to the normal weighing mode.
-------------------- At the conclusion of the “Set preload” operation, the GMP-compliant printout shown
14.01.2003 13:50 here on the left is generated.
Typ CW3P1-6ED-LCE
Ser.no. 12345678
Vers. 1.0103.11.2
BVers. 01-26-02
--------------------
Set preload
completed
--------------------
14.01.2003 13:52
Name:
--------------------
Clearing the Preload
Open the
CAL key function submenu.
The function currently set for the CAL key (in this example, “Ext. cal./adjust.; factory-
defined cal weight” (external calibration/adjustment using the weight defined at the
factory - default weight) is marked by a circle (o).
Soft key Q (repeatedly, if nec.) l q If not already selected (marked by o), activate the Clear preload setting.
To do this, press the q or Q soft key (repeatedly if necessary) to highlight this menu
Soft key l item and press the l soft key to confirm.
The “Clear Preload” item is marked by a circle (o).