89
) Adjust the scale; to do this, store linearization weight 3.
After linearization weight 3 has been stored, the target value for linearization weight 4
is shown as a negative value on the display (in this example: 6.000 kg).
§ Place the prompted weight on the scale.
After a brief pause, the difference since the most recent calibration is displayed.
$ Press ( if you wish to cancel linearization at this point.
) Adjust the scale; to do this, store linearization weight 4.
After linearization weight 4 has been stored, a zero point is prompted.
§ Remove all linearization weights from the weighing platform.
The zero point is stored automatically, after which the scale returns to the
normal weighing mode.
-------------------- When linearization has been completed, the GMP-compliant printout shown here
14.01.2003 13:00 on the left is generated.
Typ CW3P1-6DC-LCE
Ser.no. 12345678
Vers. 1.0103.11.2
BVers. 01-26-02
--------------------
Linearization
Wt.1 + 1.500 kg
Wt.2 + 3.000 kg
Wt.3 + 4.000 kg
Wt.4 + 6.000 kg
completed
--------------------
14.01.2003 13:02
Name:
--------------------