External Calibration/ Adjustment with a Factory-Defined Weight
External calibration and adjustment of scales using 100 g adjustment weight
Step Key (or instruction) Display/Printout
1. Select
SETUP in the Function line x or y
2. Confirm
SETUP u
3. Select
isoTEST 2 + y
4. Confirm
isoTEST u
5. Confirm
Weighing system settings u
6. Confirm
Calibration/adjustment u
7. Unload and tare the weighing system u
8. Select the
CAL. function y
9. Start
calibration u
32
PROGRAM
P1 105
o
C Auto.
S
0092
g
TARESETUP
0092
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
S
g
TAREEND
0000
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
S
g
TAREEND
0000
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
S
g
TAREEND
0000
KAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
J
S
g
TARAENDE