EasyManua.ls Logo

Sartorius MA150 - Page 33

Sartorius MA150
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step Key (or instruction) Display/Printout
Weighing system is being prepared
for calibration
The following is displayed when the system
is ready for calibration
10. Place the 100 g calibration weight Position the 100 g
on the weighing system (here 100.00 g) calibration weight
“–” sign: weight too low
“+” sign: weight too high
No +/- sign: weight ok
At the end of calibration the display shows
11. If no adjustment is required,
select
END in the Function line y
and confirm u --------------------
14.07.2006     15:10 
Model MA150Q-000230V 
Ser. no.    18701636 
Vers. no.   01-50-01 
ID 
-------------------- 
External calibration 
W-ID 
Nom.  +  100.000 g 
Diff. -    0.002 g 
-------------------- 
14.07.2006 15:12 
Name: 
--------------------
12. If necessary, adjust the weighing system u
At the end of adjustment, the display shows
and a report is printed: (see next page)
33
100000
PROGRAMM
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
TARASETUP
100000
PROGRAMM
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
TARASETUP
0002
END
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
ADJUSTSETUP
0002
ENDE
WGH.SYS CAL./ADJ.Verzög. 2sec
D
g
JUST.SETUP
100000
CAL.
WGH.SYS CAL./ADJ.Verzög. 2sec
S
g
TAREEND