Starting the Analysis
 –  After closing the hood:
If the initial weight condition is met, the analyzer prompts you to close the 
hood. The start weight is accepted as soon as the sample chamber is closed and 
the measurement value is stable.
End of Analysis
 – Fully automatic:
The analysis ends as soon as the weight loss per 24 s is below an automatically 
detected threshold.
 – Semi-automatic, absolute:
The analysis ends as soon as the weight loss per the set interval is lower than a 
user-defined threshold (set in milligrams). The weight loss is entered by the user.
 – Semi-automatic, percentage:
The analysis ends as soon as the weight loss in percent is lower than a user-
defined threshold (set in percent). The percentage is entered by the user.
 – Time
The analysis ends as soon as the specified time has elapsed.
 – Manual:
Switch off using the END key. The maximum analysis duration is 24 hours.
Weight
Moisture (steam)
Autom. stopping
Time