Gentle drying:
For gentle drying, you need to enter the final temperature. The maximum set
temperature is 160°C. The target temperature is reached more slowly and the
measurement time is extended. Gentle drying is recommended for sensitive samples.
Standby Temperature
Temperature set to defined value when the sample chamber is closed.
Heating only occurs when the analyzer is in a measurement cycle. The standby
temperature is not the same as the measurement temperature.
The “Timeout Standby Temp.” can be set to 1 to 24 hours and is preset to 2 hours.
To change the setting, see Chapter “System Settings (Setup Menu)”.
Initial Weight
The target weight (as a guide for applying a sample to the analyzer – not as a
starting condition) can be entered by the user. The tolerance range can be freely
selected, e.g., ± 0.5 g.
Print Lines
When print lines are activated (1 or 2 lines with 20 characters each), additional
information about the method, entered by the user, appears on the report and
printout.
Filter
The user has the option of selecting between 0, 1, or 2 filters. During taring a
message appears indicating that the configured number of filters must be placed
with the sample pan for taring.
Target Value
The user can enter a target value with a tolerance range. During analysis a progress
bar appears, indicating the progress of the measurement. At the end of the analysis,
you can see whether the value is within the tolerance range.
Standardtrocknung
T [°C]
t [s]