11
Activating the Service Mode
The LP balances are equipped with a service mode that enables expanded
calibration/adjustment functions. With these expanded functions, you do not
need to use the PSION Server or SARTOCAS Service Program for PCs to cali-
brate/adjust the balance.
The service mode only remains active until:
1. you press M to exit the menu, or
2. c is pressed when the “Service CAL Selection” menu is displayed (see
the overview on page 9).
The weighing mode is active (0.00 g is displayed - depending on the balance
model)
Enter the service password: 202122.
Press M.
The SETUP menu is displayed and an “S” is shown in the lower left corner.
The service mode is now active.
Press c to exit the SETUP menu; press q to activate the calibration/
adjustment menu.
Calibration/Adjustment in the Service Mode
Adjustment options in the service mode:
1. int. adjust * Span adjustment with internal weight
2. def. ext.lin. External linearization with default weight
3. ext. linear. External linearization (enter weight value in the IN-
PUT mode)
4. reprotest * Reproducibility test of the internal adjustment
5. det. int. wt. Overwrite the internal calibration weight
6. det. ext. wt. Determine and store the external weight (user-defined
weight)
7. disp. *10 10-fold increased display resolution; SuperRange
resolution
8. def. ext.adj. * External span adjustment with default weight
9. auto ext.adj.* External span adjustment with automatic recognition
of calibration weight
10. ext.adj. * External span adjustment with stored (user-defined)
weight (entered in the INPUT mode)
Note: In most cases it is not necessary to use service software programs (PSION or
Sartocas) to adjust the LP balances, because all necessary adjustment func-
tions are accessible when the balance is in the service mode.
You only need to use the adjustment software if the current adjustment error is
greater than 2% or if multiple adjustment steps are necessary due to unfavor-
able ambient conditions.
0.00 g
Max5200 g d= 0.01 g
202122
S-ID
Max5200 g d= 0.01 g
202122
Sp. ID
S INFO MENU INPUT
*
Can also be adjusted without activating the service mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Master Pro LA5200P and is the answer not in the manual?

Sartorius Master Pro LA5200P Specifications

General IconGeneral
BrandSartorius
ModelMaster Pro LA5200P
CategoryScales
LanguageEnglish

Related product manuals