Compare the voltage displayed on the digital voltmeter with the pre-load
voltage given in the „Adjustment Specifications.“ If necessary, switch the
voltmeter to a lower measuring range (e.g., 200 mV DC).
If the pre-load voltage measured exceeds the tolerance limits listed in the
specifications, you must adjust the preload.
The data for preload test weights and preload voltage are listed in the
“Adjustment Specifications” below.
Adjustment Specifications
LA/LP balances with brass system:
LA/LP1200S, LA/LP3200D, LA/LP4200S, LA/LP5200P, LA/LP6200S
Models Preload test weight Preload voltage A/D converter
level control
LA/LP1200S 600 g –400 mV to +400 mV 20 % ± 2.3 %
LA/LP3200D 1500 g –260 mV to +260 mV 20 % ± 1.5 %
LA/LP4200S 300 g –270 mV to +270 mV 20 % ± 1.6 %
LA/LP5200P 300 g –270 mV to +270 mV 20 % ± 1.6 %
LA/LP6200S 300 g –270 mV to +270 mV 20 % ± 1.6 %
LA/LP balances with small single-casting block system:
LA/LP420, LA/LP220S, LA/LP620S, LA/LP620P, LA/LP820, LA/LP2200,
LA/LP2200S, LA/LP2200P, LA/LP4200, LA/LP6200, LA/LP12000S, LA/
LP12000P
Models Preload test weight Preload voltage A/D converter
level control
LA/LP220S 300 g –360 mV to +360 mV 20 % ± 1.2 %
LA/LP420S 2150 g –590 mV to +590 mV 20 % ± 3.4 %
LA/LP620S 300 g –360 mV to +360 mV 20 % ± 2.1 %
LA/LP620P 300 g –360 mV to +360 mV 20 % ± 2.1 %
LA/LP820 300 g –360 mV to +360 mV 20 % ± 3.4 %
LA/LP2200 2150 g –590 mV to +590 mV 20 % ± 3.4 %
LA/LP2200S 2150 g –590 mV to +590 mV 20 % ± 3.4 %
LA/LP2200P 2150 g –590 mV to +590 mV 20 % ± 3.4 %
LA/LP4200 6000 g –200 mV to +200 mV 20 % ± 1.2 %
LA/LP6200 6000 g –200 mV to +200 mV 20 % ± 1.2 %
LA/LP12000S 3000 g –200 mV to +200 mV 20 % ± 1.2 %
LA/LP12000P 6000 g –200 mV to +200 mV 20 % ± 1.2 %
LA/LP balances with large single-casting block system:
LA/LP16000S, LA/LP34000P, LA/LP34
Models Preload test weight Preload voltage A/D converter
level control
LA/LP16000S 17 kg –130 mV to +130 mV 20 % ± 0.75 %
LA/LP34000P 17 kg –130 mV to +130 mV 20 % ± 0.75 %
LA/LP34 17 kg –130 mV to +130 mV 20 % ± 0.75 %
Solderside
A
B
Leitpl1a.EPS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Master Pro LA5200P and is the answer not in the manual?

Sartorius Master Pro LA5200P Specifications

General IconGeneral
BrandSartorius
ModelMaster Pro LA5200P
CategoryScales
LanguageEnglish

Related product manuals