When data are output without decimals, the decimal point is suppressed
(except when a certain display mode is selected).
12345678910111213141516
Data output example: +1255.7 g
12345678910111213141516
Characters:
1st Plus or minus sign or space
2nd Space
3rd–10th Weight with a decimal point; leading zeros = space
11th Space
12th–14th Unit symbol or space
15th Carriage return (CR)
16th Line feed (LF)
If the weighing system has not stabilized, no unit symbol will be output.
Unit symbols:
Note:
When the scale is used as a legal measuring instrument, all units except g/kg
are blocked.
***
No stability parameter t l t Taiwanese taels
o
**
Grams (o) G N
*
Grains
g
**
Grams d w t Pennyweights
kg
*
Kilograms / l b Parts per Pound
ct
*
Carats t l c Chinese taels
lb
*
Pounds m o m Mommes
oz
*
Ounces K
**
Austrian carats
o z t Troy ounces t o l Tola
t l h Hong Kong taels b a t Baht
t l s Singapore taels MS
*
Mesghal
* = space, U = Unit
+
***
1255 . 7
*
g
**
CR LF
+
******
–––––––––––––––––––––
***
***
10
6
10
5
10
4
10
3
10
2
10
1
10
0
*
CR LF
––––––––––––––––––––– U U U
000000