Calibration/Adjustment and Linearization Functions Using the =/z Key
(hold down key for more than 2 seconds)
Important Note:
For verifiable scales used as legal measuring instruments the menu access switch is
blocked and sealed. The external calibration is not permitted.
External Calibration and/or Linearization
The weights to be placed on the scale for linearization are indicated in the display
one after the other, in increasing order.
Calibration/adjustment and linearization Code Changes
External calibration 1 9 1 *
External linearization 1 9 5
Adjustment/calibration function blocked 1 9 7 **
* = factory setting
** = blocked for all veryfiable models