Scale Operating Parameters
Adapting the Scale to Ambient Conditions
The scale can be adapted to the prevailing ambient conditions at the place
of installation.
Code Changes
Very stable conditions 1 1 1
Stable conditions 1 1 2 *
Unstable conditions 1 1 3
Very unstable conditions 1 1 4
Standard Weighing Mode – Manual Filling Mode
You can optimally adapt your scale to meet either of these requirements.
In the manual filling mode, the display compensates for fluctuations of the load
on the scale, giving you especially fast and stable readouts.
Code Changes
Standard weighing mode 1 2 1 *
Manual filling mode 1 2 2
(e.g. if you need to package screws in certain quantities)
Stability Range
The stability symbol will remain displayed in the case of a weight variation +/–.
Code Changes
0.25 digit 1 3 1
0.5 digit 1 3 2 ***
1 digit
1
) 133
2 digits
1
) 1 3 4 * (**)
4 digits
2
) 135
8 digits
2
) 136
* Settings for standard models
** Settings for QC 7CCE-S0CE and QC 64EDE-S0CE models
*** Settings for QC 7-,34-,64- L0CE and QC 34EDE-S0CE models
1
) blocked for all verifiable models of accuracy class m.
2
) blocked for all verifiable models
2–4