Utilities for Printouts or Data Transfer
Sartorius QC scales come standard with an interface.
You can plug a Sartorius printer or a computer into this interface port. In addition,
you can choose to have data output from your scale to this on-line device either
automatically or by pressing the p/p key.
The scale operating menu lets you define the various parameters for data output.
For information on the data formats and for interfacing a computer or a different
peripheral device, see Part 4, “General Description of the Data Interface.”
Data Output Parameter
This parameter is coupled with the stability parameter; stability = stable readout
or no motion is detected.
Print on request = data is output only when the p/p key is pressed
or a software command is received
Auto print = continuous, automatic data output
Code Changes
Print on request regardless of stability 6 1 1
Print on request after stability, with storage of the function 6 1 2 *
Print on request after stability, without storage of the function 6 1 3
Auto print regardless of stability 6 1 4
Auto print at stability 6 1 5
Auto Print
You can stop and start automatic data output (auto print function) by pressing
the p/p key (11). To avoid operating errors or to ensure that the data will be
output continuously in the automatic mode, you can block this function.
Start/stop auto print using the Code Changes
p/p key 6 2 1
Auto print not stoppable 6 2 2 *
* = factory setting
2–9