Data Output with ID Code
When data with an ID code is output, the ID code consisting of 6 characters
precedes the data with the16-character format.
During data output, all characters are shifted to the right by 6 places.
1st
7th 22nd
character
V = Plus or minus sign
* = Space
x = Digit
U = Unit
. = Decimal point
K = Letter for an ID comment
CR = Carriage Return
LF = Line Feed
When special codes are output, the letters “Stat” for status code are assigned to the
1st through the 4th characters of the data string.
Status-dependent string:
1st
7th 13th14th 22nd
character
A, B = status codes
Error-dependent string:
1st
7th 10th
12th 14th
16th 22nd
character
Stat
*****
ERR
*
XYZ
****
CR LF
Stat
********
AB
******
CR LF
C C C C C C V
*
x x x x x x x x
*
U U U CR LF
* * * * * * * * . . . . . . * * *