Step/Key Display Data Output
The user guide display shows
S– – = d – min. weight +
200 pcs
Enter the tolerance value
(e.g.1piece)
using the numeric keys 1
0 gcs
1 gcs
Press the z/T key again +
100.1 pcs
The user guide display shows
S–t = d – max. weight +
200 pcs
Enter the tolerance value
(e.g.1piece) using the
numeric keys 1
0 gcs
+
1gcs
Press the z/T key again +
200 pcs
An acoustic signal is given
and data output is generated
automatically. Setp + 200 pcs
d-min - 1 pcs
d-max + 1 pcs
Diff + 0 pcs
GOOD values are
output automatically.
If the package is too heavy,
this is indicated by HH HH
in the display. If the weighed
value is within the GOOD
range, output is generated
automatically and an
acoustic signal is given.
Note:
A number entered manually without a decimal point is always a piece count!
A number entered manually with a decimal point is always a weight value!