Testing the Monitor 35
SCOTT SAFETY 087-0065 REV B
Bump Test This section describes how to perform a manual Bump Test.
1 Refer to “Connecting the Gas Cylinder to Monitor” on page 33 to install the
calibration kit.
2 Press and hold the left button, a screen appears stating Bump Test.
3 Apply gas and attach Calibration Adapter to monitor (Figure 14).
4 Press the left button to acknowledge the bump test and it will apply gas.
5 Verify monitor responds to target gas and wait for the display to read PASS. Verify
the device activates 1 red LED flash per second, 1 sounding beep per second and 1
vibration per second. Press any button or wait 60 seconds to go to the normal
reading screen.
6 After an unsuccessful bump test, the monitor LCD reads FAIL. Verify the device
activates 1 red LED flash per second, 1 sounding beep per second and 1 vibration
per second. Press any button or wait 60 seconds to go to the normal reading screen.
7 Turn off gas cylinder and remove Calibration Adapter.