26 CHAPTER 4: OPERATION
087-0065 REV B SCOTT SAFETY
Monitor LCD Alerts &
Alarms
This section describes various alerts and alarms.Table 14 lists the details.
To view the factory default alarm setpoints, refer to Default Settings” on page 17.
To change the default settings and values, refer to Portable Gas Detection (PGD)
Application User Manual (087-0061).
Table 14 Monitor Alerts and Alarms Descriptions
LCD DESCRIPTION LCD DESCRIPTION
Over Limit Exposure Alarm
3 fast LED flashes every 2
seconds
3 fast sounding beeps every 2
seconds
1 vibration every 2 seconds
“++++” appears
Negative Drift Alarm
1 slow LED flash every 2
seconds
1 slow sounding beep every 2
seconds
1 vibration every 2 seconds
and gas reading flashes
High Alarm
3 fast LED flashes every 2
seconds
3 fast sounding beeps every 2
seconds
1 vibration every 2 seconds
HIGH and gas reading flashes
Low Alarm
2 fast LED flashes every 2
seconds
2 fast sounding beeps every 2
seconds
1 vibration every 2 seconds
LOW and gas reading flashes
TWA Alarm
1 LED flash every 2 seconds
1 sounding beep every 2
seconds
1 vibration every 2 seconds
TWA and gas reading flashes
STEL Alarm
1 LED flash every 2 seconds
1 sounding beep every 2
seconds
1 vibration every 2 seconds
STEL and gas reading flashes
Low Battery Alarm
Battery icon flashes
1 Beep, 1 LED flash and 1
vibration every minute
The device deactivates with
the countdown (depleted)
Bump Test Due
BUMP flashes
DUE flashes
O2
O2
%
HIGH
O2
%
LOW
H2 S
PPM
TWA
H2 S
PPM
STEL
CO
PPM

Table of Contents