Figure 4-6. Bump Test Setup
6) Verify Protégé quickly responds to target gas by activating the visual, audible, and vibrating
alarms.
7) Verify readings on the LCD display within 10% of actual target gas concentration.
8) Turn off gas cylinder and remove calibration adapter.
WARNING
IF THE PROTÉGÉ FAILS TO ACTIVATE ALL ALARMS IN A TIMELY MANNER, THE UNIT MUST BE SERVICED PRIOR TO
USAGE. IF THE PROTÉGÉ FAILS TO READ TARGET GAS CONCENTRATION WITHIN 10%, PERFORM A SPAN
CALIBRATION. FAILURE TO DO SO COULD LEAD TO DEATH OR INJURY.