33
5. Calibration Procedure
5.1 Calibration mode start method
1) Turn on the PC with the maintenance tool installed (hereinafter referred to as PC).
2) Log on as necessary.
3) Start the maintenance tool.
Select and click “
(Start)” - “Programs” - “SEKONIC Corporation” – “L-478 Maintenance Tool”.
4) Connect PC and L-478 to update with USB cable, and turn on the L-478.
5) Depending on a kind of the adjustment, click the following buttons
Calibration Mode Button to click
Battery Calibration
Set to the Battery calibrationButton()
Incident light calibration
Set to the Incident light calibrationButton ()
touch panel Calibration
Set to the Touch panel calibrationButton ()
Cancel the calibration modeButton() is clicked when it starts in adjustment mode by mistake.
6) Turn off L-478
7) To end the program, click the “Exit” button.
8) The specified adjustment mode starts at the time of next L-478 power-supply ON.