4
1. Equipment, tools and materials required for repair
1. Calibration box
The attachment to light box for L-478 incident light(for L-478 only)
2. Master exposure meter
3. Digital multi-meter (capable of measuring voltage, current and resistance)
4. Power supply box or Constant voltage power supply (capable of varying voltage, current and resistance)
(for L-478 only)
Calibration voltage: 2.4V+/-0.01V、3.0V+/-0.01V
Confirmation voltage:2.1V+/-0.01V、2.23V+/-0.01V、2.33V+/-0.01V、2.4V+/-0.01V、2.6V+/-0.01V
*1 All voltage should be the terminal voltage values at the time of exposure meter ON.
*2 Current capacity : 0.5A and more
5.Torque Driver : Phillips(+) /#1, general tools,
Dummy battery(for L-478 only)
6.
IC-use soldering iron and solder (Pb Free Wire Solder Sn3.0Ag0.5Cu)
M705 SENJU METAL INDUSTRY Co., Ltd., or equivalent.
7. Blower brush, Nonwoven cloth
8.
Gloves
9. Adhesive : 1) DB1600B Diabond Industry Co., Ltd., product or equivalent
: 2) Super-X No.8008,CEMEDINE CO., LTD. product or equivalent
:3) Type KE-45_1 RTV Rubber, Shin-Etsu Chemical Co., Ltd product or equivalent.
10. Grease : G501 Shin-Etsu Chemical Co., Ltd., product or equivalent
11. Screw lock : 1401B Three Bond Co., Ltd., product or equivalent
12. Windows PC(for L-758 Maintenance Control)
Required OS :
Microsoft Windows XP Service Pack3 or Later.
or Microsoft Windows VISTA Service Pack2 or Later.
or Microsoft Windows 7 Service Pack1 or Later.
Recommended OS :
Microsoft Windows 7 Service Pack1 or Later.
Required Memory :
256MB (Recommended 512MB)
Required Hard disk :
100MB
Remarks : Please warn security enough
13. CD supplied by SEKONIC
L-478 Maintenance Tools
14. USB Cable (JH109001 Forward Wire Harness or equivalent: A-MiniB type)