37
10) "TP calibration end" is displayed on middle of the screen.
11)Press the Power button (on/off switch) and turn the power supply off.
5.5 Errors during calibration
Error is displayed when applied to the following conditions at the time of Battery Calibration and Incident light
calibration.
Error Type Criteria
Err1
1
Variations in battery correction count exceeds ±5 of the mean
2
Variations in Incident light correction count exceeds ±2 of the mean
Err2
1
Variations in temperature count during Incident light correction exceeds ±20 of the mean
Err3
1
Correction is performed with the limits of correction value inverted
2
Correction is performed with the same input value
Error Type Countermeasure
Err1
1
Press the measurement button. Measurement is performed immediately after error is
released.
Err2
Err3
Error is displayed when applied to the following conditions at the time of touch panel Calibration
Error Type Criteria
Screen blink
1
When the position which exceeded ±15 dots (an X coordinate or a y coordinate) from the
center of the cross mark is touched in "5.4 touch panel calibration procedure 6 - 9."
Error Type Countermeasure
Screen blink
1
Since a cross mark is re-displayed, touch the right position.
2
Push a memory button.
Redo can be performed from "5.4 touch panel calibration procedure 2”.