20
3.2 Methods for Device Repair, Adjustment, etc.
3.2.1 JR10-616 Optics Assembly / 1 (lumisphere) related malfunctions. (Foundations are unit
exchange).
In the event of lens movement malfunctions, such as not moving vertically or catching, disassemble
the relevant component while referring to the structural diagram, confirm part damage, deformation or
the presence of foreign matter (dust, etc.), replace any parts that are damaged or deformed,
Cautions during assembly
・ZA356 : The screw tightening torque is to be 3~5cNm (2.9~4.9kgcm)
*A screw lock (Three Bond 1401B) is applied at the tip of a screw.
・JR106106 : The screw tightening torque is to be 3~5cNm (2.9~4.9kgcm)
*A screw lock (Three Bond 1401B) is applied at the tip of a screw.
・When you attach a Cam Case Assembly (JR10-617) to a Optics Guide Case Assembly (JR10-618),
be careful of fall of a steel ball (ZT20 : 2 mm).
*A steel ball will become difficult to fall if silicon grease (G501) is applied to a steel ball.
・When you attach a Cam Case Assembly (JR10-617) to a Optics Guide Case Assembly (JR10-618) ,
be careful in the attachment direction.
・Please apply silicon grease (G501) to the slot (2pcs) of a Optics Guide Case Assembly (JR10-618) .
JR10-618
Optics Guide Case Assembly
Slot : 2pcs
JR10-619
Lens Fixing Assembly
ZA356 (M1.4×3)
2pcs
JR106107
Cam Fixing Plate
JR10-618
Optics Guide Case Assembly
ZT20
Steel Ball (2mm) 2pcs
JR10-617
Cam Case Assembly
JC107411
Spring / 1 2pcs
JR106106
Cam Screw
2
sc
Please unite the slit of a Optic
Guide Case Assembly, and th
slit of a Cam Case Assembly.