36
5.4 touch panel Calibration
Required tools: PC, USB cable, battery (Size AAA battery), Touch pen
1) A battery is put into meter. And In accordance with "5.1 Calibration mode start method", it starts in
touch panel
Calibration mode.
2) Since a cross mark is displayed on the upper left, touch a center with a touch pen.
3) Since a cross mark is displayed on the lower right, touch a center with a touch pen.
4) Since a cross mark is displayed on the upper right, touch a center with a touch pen.
5) Since a cross mark is displayed on the lower left, touch a center with a touch pen.
6) Since a cross mark is displayed on the upper left, touch a center with a touch pen.
Since a red cross mark is displayed, check that it is in the almost same position as a white cross
mark.
7) Since a cross mark is displayed on the lower right, touch a center with a touch pen.
Since a red cross mark is displayed, check that it is in the almost same position as a white cross
mark.
8) Since a cross mark is displayed on the upper right, touch a center with a touch pen.
Since a red cross mark is displayed, check that it is in the almost same position as a white cross
mark.
9) Since a cross mark is displayed on the lower left, touch a center with a touch pen.
Since a red cross mark is displayed, check that it is in the almost same position as a white cross
mark.