1 Introduction
HB-005615-j
6
WARNING: The SDMS only can be used in patients undergoing
hyperbaric therapy if the SDM (monitor) remains outside the
hyperbaric environment.
WARNING: The SDM is protected against electrostatic/
defibrillator discharge. Parameter display may be temporarily
affected during discharge/ defibrillation, but will rapidly recover.
Precisely follow the instructions given in the defibrillator manual.
WARNING: This device has been tested and found to comply
with the requirements for medical devices according to the IEC
60601-1-2, and the Medical Device Directive 93/42/EEC. These
requirements are designed to provide reasonable protection
against harmful interference in a typical medical installation.
WARNING: To ensure protection of patient, operator and
equipment from the effects of defibrillation and diathermy/
electro-surgery, cables manufactured by SenTec AG must be
used.
WARNING: During normal operation (except transport), it is
recommended that the monitor is always connected to the AC
power outlet.
WARNING: Do not connect the SDM to an electrical outlet
controlled by a wall switch, because the SDM may be accidentally
turned off once the battery is depleted.
WARNING: To avoid risk of electric shock, this equipment
should be connected to a supply mains with protective earth.
Ensure that power and protective ground lines are connected
correctly. If case of doubt (e.g. as this may be the case during
home use of the SDM) disconnect the SDM from the outlet and
use the battery power during patient monitoring.
WARNING: Do not use the protective earth terminal for any
other purpose than protective earthing or functional earthing.
WARNING: For US, respectively Japan: Grounding reliability
can only be achieved when the SDM is connected to an
equivalent receptacle marked HG (Hospital Grade), respectively
HGJ (Hospital Grade Japan).