Verify that the sensor is securely attached
and ensure that air gaps are eliminated
between the skin and the sensor. If needed,
reapply the sensor (on same or an alternate
site). For instructions refer to the
Quick
Reference Guide for the SDMS
or the
Instruction Manual for the SDMS
.
Consider to use the Staysite™ Adhesive
(model SA-MAR) to complementary attach
Multi-Site Attachment Rings if more secure
sensor attachment is required, e.g. in high
humidity environments, for patients who
perspire profusely and/or in challenging
patient motion conditions.
Note: Typically PaCO2 is > 25 mmHg in
humans and PCO2 of ambient air < 0.5
mmHg. If the contact between the sensor
and skin is not hermetically sealed PCO2
readings, consequently, may increase slowly
after sensor application or a ‘TC artifact’ and
be too low thereafter.
Note: Typically PaO2 is < 100 mmHg in
humans and PO2 of ambient air > 130
mmHg. If the contact between the sensor
and skin is not hermetically sealed PO2
readings, consequently, may decrease slowly
after sensor application or a ‘TC artifact’ and
be too high thereafter.
Note: Also refer to Status Messages ‘Check
sensor application’, ‘PCO2 stabilizing’, ‘PO2
stabilizing’, or ‘PCO2/ PO2 stabilizing’ (refer to
the
Technical Manual for the SDM
).