8.8 SDMS Maintenance Checklist
This checklist is intended to be used by the operator.
Please refer to the document:
HB-007601
8.9 Interfaces - Pin Assignement
WARNING: The mains power supply of the SenTec Digital
Monitor (SDM) is separated by two Means of Patient Protection
(MOPPs) between the sensor port (for the applied part, the
sensor) and the interface connectors. The three interface
connectors - serial data port, Multipurpose I/O port (analog
outputs, nurse call), LAN port - of the SDM are not separated
from each other. If at a time accessory equipment is connected
to only one of the three interface connectors no additional safety
measures are necessary to comply with the requirements of IEC
60601-1. If however accessory equipment is simultaneously
connected to two or three of the SDM’s interface connectors
additional safety measures may be required to be compliant with
the requirements of IEC 60601-1. In case of doubt consult
qualified technicians.