3 Maintenance/Safety & Functionality Tests
HB-005615-j
30
3.2 SenTec Digital Monitor
3.2.1 Cleaning and Disinfection
Clean or disinfect the SDM (including AC power cord) as often as required per
institutional ordinances (recommended weekly). For cleaning and disinfection
instructions refer to the
Technical Manual for the SDM
(HB-005752).
3.2.2 Visual Inspection
Visually check the entire SDM for possible damages.
If you detect a damage, the SDM needs to be repaired according to
section 5 "Repairs".
3.2.3 Labeling
Check the SDM for presence and legibility of the following labels and symbols:
Front panel: Symbols for buttons and LED’s (
see Figure 1 Front Panel of the SDM
).
Rear panel: Labels of all ports, connectors and switches; label with SN, REF and
, CE label, UL label and WEEE Disposable label (
see Figure 2 Rear
Panel of the SDM
).
If the above requirements are not met contact your local SenTec representative or
send back the monitor according to
section 8.6 "Instructions for Shipment.
For SenTec authorized service technicians only: If the above requirements
for the front panel are not fulfilled then replace the SDM Case, upper part according
to the
Repair Manual for the SDMS
(HB-005755).
3.2.4 Electrical Safety Check
The SenTec Digital Monitoring System (SDMS) has been designed to comply with the
standards IEC 60601-1, ANSI/AAMI ES60601-1 and CAN/CSA-C22.2 No.60601-1 (UL
and cUL). To ensure electrical safety of the equipment it is recommended that
qualified service personnel performs the electrical safety check once a year (but
every two years the latest) and after repair.
The electrical safety tests for the SDMS have to be performed in accordance with the
standards IEC 60601-1, ANSI/AAMI ES60601-1 (USA) or CAN/CSA-C22.2 No.60601-1
(Canada) for instruments classified as Class 1 and Type BF (defibrillation proof).
Technicians must be familiar with these standards applicable to the institution and the
country. Test equipment and its application must comply with the applicable standard.