EasyManua.ls Logo

Sentec SDM - Page 27

Sentec SDM
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Maintenance/Safety & Functionality Tests
SenTec Digital Monitoring System
21
6. Insert the test bottle with 5% CO2 and 6% O2 into the Sensor Tester. Use the
menu-function PCO2 Settings / Calibrate Sensor or press the ENTER Button
twice to start a calibration. Record PCO2 (5%) and PO2 (6%), i.e. the PCO2/PO2
values just before termination of the calibration. After the calibration is finished
please wait 6 minutes (this is to ensure that no gas leak is present in the Docking
Station of the SenTec Sensor Tester).
Note: If you are using an SDM for the Slope Test and in case of an OxiVenT™
Sensor the message `Sensor Problem 73´ appears on the SDM at this stage. This
is an expected behavior of the sensor.
7. Compare the measured PCO2 (5%) and PO2 (6%) with the values expected for
the respective local atmospheric pressure (see Slope Test Control Table). PCO2
(5%) should be within 5_min and 5_max. PO2 (6%) should be within 6_min and
6_max.
8. Remove the sensor from the Docking Station of the Sensor Tester, insert the
sensor into the Docking Station of the SDM and use the menu-‘PCO2 Settings /
Calibrate Sensor’ to start a calibration (with 8% CO2, 12% O2).
Note: In case of an OxiVenT™ Sensor the message `Sensor Problem 73´ will
disappear from the SDM`s display at this stage.
9. Insert the test bottle with 30% CO2 and 21% O2 into the Sensor Tester. Use the
menu-function ‘PCO2 Settings / Calibrate Sensor’ or press the ENTER Button
twice to start a calibration. Wait until calibration is finished and record PCO2
(30%) and PO2 (21%), i.e. the PCO2 value just before termination of the
calibration.
Note: If you are using an SDM for the Slope Test and in case of an OxiVenT™
Sensor the message `Sensor Problem 73´ appears on the SDM at this stage. This
is an expected behavior of the sensor.
10. Compare the measured PCO2 (30%) and PO2 (21%) with the values expected
for the respective local atmospheric pressure (see Slope Test Control Table).
PCO2 (30%) should be within 30_min and 30_max. PO2 (21%) should be within
21_min and 21_max.
11. In case of an OxiVenT™ Sensor: In order to clear the message `Sensor Problem
73´ you must do either of the following: a) power-cycle the SDM; b) connect the
OxiVenT™ Sensor to a different SDM; or c) perform a calibration with the
standard gas concentration (i.e. with 8% CO2, 12% O2).
12. If you have used one single SDM for the Slope Test, always perform a calibration
with the standard gas concentration (i.e. with 8% CO2, 12% O2) at the end of
the Slope Test.
If the readings are outside the respective limits replace the sensor and
report to SenTec the sensor behavior/replacement.

Table of Contents

Other manuals for Sentec SDM