3 Maintenance/Safety & Functionality Tests
HB-005615-j
26
Slope Test Control Table PO2 in kPa (after 6% and 21% O2 calibration)
Pcal = value after calibration with 12% O2
Table is only valid for sensor temp. of 42°C.
Patm
Pcal
6_min
6_ideal
6_max
21_min
21_ideal
21_max
80.0
9.6
4.6
4.8
5.0
15.8
16.8
17.8
80.7
9.7
4.6
4.8
5.1
15.9
16.9
18.0
81.3
9.8
4.6
4.9
5.1
16.0
17.1
18.1
82.0
9.8
4.7
4.9
5.2
16.2
17.2
18.3
82.7
9.9
4.7
5.0
5.2
16.3
17.4
18.4
83.3
10.0
4.7
5.0
5.2
16.4
17.5
18.5
84.0
10.1
4.8
5.0
5.3
16.6
17.6
18.7
84.7
10.2
4.8
5.1
5.3
16.7
17.8
18.9
85.3
10.2
4.9
5.1
5.4
16.8
17.9
19.0
86.0
10.3
4.9
5.2
5.4
17.0
18.1
19.1
86.7
10.4
4.9
5.2
5.5
17.1
18.2
19.3
87.3
10.5
5.0
5.2
5.5
17.2
18.3
19.4
88.0
10.6
5.0
5.3
5.5
17.4
18.5
19.6
88.7
10.6
5.1
5.3
5.6
17.5
18.6
19.7
89.3
10.7
5.1
5.4
5.6
17.6
18.8
19.9
90.0
10.8
5.1
5.4
5.7
17.8
18.9
20.0
90.7
10.9
5.2
5.4
5.7
17.9
19.0
20.2
91.3
11.0
5.2
5.5
5.8
18.0
19.2
20.3
92.0
11.0
5.2
5.5
5.8
18.2
19.3
20.5
92.7
11.1
5.3
5.6
5.8
18.3
19.5
20.6
93.3
11.2
5.3
5.6
5.9
18.4
19.6
20.8
94.0
11.3
5.4
5.6
5.9
18.6
19.7
20.9
94.7
11.4
5.4
5.7
6.0
18.7
19.9
21.1
95.3
11.4
5.4
5.7
6.0
18.8
20.0
21.2
96.0
11.5
5.5
5.8
6.0
19.0
20.2
21.4
96.7
11.6
5.5
5.8
6.1
19.1
20.3
21.5
97.3
11.7
5.5
5.8
6.1
19.2
20.4
21.7
98.0
11.8
5.6
5.9
6.2
19.3
20.6
21.8
98.7
11.8
5.6
5.9
6.2
19.5
20.7
22.0
99.3
11.9
5.7
6.0
6.3
19.6
20.9
22.1
100.0
12.0
5.7
6.0
6.3
19.7
21.0
22.3
100.7
12.1
5.7
6.0
6.3
19.9
21.1
22.4
101.3
12.2
5.8
6.1
6.4
20.0
21.3
22.5
102.0
12.2
5.8
6.1
6.4
20.1
21.4
22.7
102.7
12.3
5.9
6.2
6.5
20.3
21.6
22.9