4 Troubleshooting
SenTec Digital Monitoring System
SMBM
(Motherboard)
defective
Replace SMBM according to the
Repair
Manual for the SDMS
.
To be performed by SenTec authorized
service technicians only.
P0416
No alarm on
speaker,
nurse call
Auditory alarm signals can be muted for 1
minute, 2 minutes or permanently with the
Alarm Mute Button.
Push the Alarm Mute Button.
Refer to P0411.
Contact qualified service personnel.
Contact SenTec authorized service personnel.
SMBC (Controller
Board) defective
Replace SMBC according to the
Repair Manual
for the SDMS
.
To be performed by SenTec authorized
service technicians only.
SMBM
(Motherboard)
defective
Replace SMBM according to the
Repair
Manual for the SDMS
.
To be performed by SenTec authorized
service technicians only.
P0417
Safety test
does not
comply
Contact SenTec authorized service personnel.
Protective Earth
Resistance is more
than 0.2 Ω
Control the Protective Earth wiring between
the AC Power Connector, the Equipotential
Terminal and the Gas bottle.
To be performed by SenTec authorized
service technicians only.
Protective Earth
wiring defective
If problem persists, replace Protective Earth
according to the
Repair Manual for the SDMS
.
To be performed by SenTec authorized
service technicians only.
Replace SMBM according to the
Repair
Manual for the SDMS
.
To be performed by SenTec authorized
service technicians only.