SERVOPRO NOx Analyzers Page
On/Off:
0 = Off
1 = On
AVER: Query Software Version
Command Response Description
_AVER_K0 _AVER_s_NMAIN_Z_mm.dd.yyyy_
NUSER_Y_mm.dd.yyyy_OSMSR_
X_dd.mm.yyyy
Z: NMAIN version build number
Y: NUSER version build number
X: OSMSR version build number
dd: Day
mm: Month
yyyy: Year
CONTROL COMMANDS
SARA: Auto Range Off
Command Response Description
_SARA_K0 _SARA_s Sets auto range off (CLD and FID)
SARE: Auto Range On
Command Response Description
_SARE_K0 _SARE_s Sets auto range on
SATK: Start Automatic Calibration
Command Response Description
_SATK_K0 _SATK_s Starts automatic calibration of all available ranges
_ SATK_K0_Mn _SATK_s Starts automatic calibration of range n
SEGA: Open Valve for Span Gas Calibration