SERVOPRO NOx Analyzers Page
Command Response Description
_SEGA_K0 _SNGA_s Sets NO to span calibration mode
_ SEGA_K1 _SNGA_s Sets O
2
to span calibration mode
_ SEGA_K0_Mn _SNGA_s Opens NO to span calibration mode Range n
SEKA: Saves Measured Value as New Span Value
Command Response Description
_SEKA_K0 _SEKA_s Saves measured value of actual range as gain if
Span mode is active
_ SEKA_K1 _SEKA_s Saves measured value as gain of O
2
channel if O
2
Span mode is active
SEMB: Set Measuring Range
Command Response Description
_SEMB_K0_Mn _SEMB_s Sets measuring range to range n. Auto range
will be disabled
SENT: Set Calibration Gas Flow (Pumps or Valves)
Command Response Description
_SENT_K0_X _SENT_s x: 10 = Pump
11 = Valves
SFGR: Reset Calibration Settings to factory defaults
Command Response Description
_SFGR_K0 _SFGR_s Resets all calibration settings to their factory
settings
SMAN: Manual Control to Control Device Manually
Command Response Description