SERVOPRO NOx Analyzers Page
_SMAN_K0 _SMAN_s Sets analyzer in manual mode
SMGA: Start Measuring; Turn On Pumps if Fitted
Command Response Description
_SMGA_K0 _SMGA_s Sets analyzer to measure mode
SNGA: Open Valve for Zero Gas Calibration
Command Response Description
_SNGA_K0 _SNGA_s Sets NO to zero calibration mode
_ SNGA_K1 _SNGA_s Sets O
2
zero calibration mode
_ SNGA_K0_Mn _SNGA_s Sets NO to zero calibration mode for Range n
SNKA: Saves Measured Value as New Offset
Command Response Description
_SNKA_K0 _SNKA_s Saves measured value of actual range as offset if
Zero mode is active
_ SNKA_K1 _SNKA_s Saves measured value as offset of O
2
channel if
O
2
Zero mode is active
SPAU: Pause
Command Response Description
_SPAU_K0 _SPAU_s Sets Pause mode
SREM: Remote Mode for AK Commands
Command Response Description
_SREM_K0 _SREM_s Sets analyzer in Remote mode
SRES: Reset
Command Response Description
_SRES_K0 _SRES_s Reset