SERVOPRO NOx Analyzers Page
SSPL: Purge Analyzer with Zero Gas
Command Response Description
_SSPL_K0 _SSPL_s Opens purge gas valve
STBY: Standby
Command Response Description
_STBY_K0 _STBY_s Standby mode
SUDP: Start/Stop UDP Data Streaming
Command Response Description
_SUDP_K0_ON _SUDP_s Starts data streaming via the UDP channel.
Configure the UDP channel before starting with
EUDP command.
_ SUDP_K0_OFF _SUDP_s Stops streaming via the UDP channel
SVZS: Reset Offset to 0 and Gain to 1
Command Response Description
_SVZS_K0 _SVZS_s Sets all range offsets to 0 and all gains to 1
_ SVZS_K1 _SVZS_s Sets O
2
range offsets to 0 and gains to 1
CONFIGURATION COMMANDS
EATK: Set Auto Calibration Parameters
Command Response Description
_EATK_K0_z_y_x _EATK_s_z_y_x z: 1) NO mode
2) NO
x
mode
y: 1) ALL gases