2.3
WARNING
C The Xentra 4900 is not suitable for use in hazardous areas.
C The Xentra 4900 is not suitable for use with corrosive or
flammable samples.
C The sample gas may be toxic or asphyxiant and must be
vented to a safe location.
CAUTION
Install the 4900 so that fan and cover vents are not obstructed.
2.3 Panel mount installation
See Figure 2.2 for panel mounting detail. In panel mounting format the xentra 4900
is supplied with a pair of mounting brackets ( item 1 on figure 2.2 ) suitable for mounting
the front of the instrument against a panel. These brackets are not intended to provide
the sole means of support for the analyser. The user must provide additional means
of support at the rear of the instrument ( item 2 on figure 2.2 ).
WARNING
The rack mounting brackets are not intended to provide the sole means
of support. The user must provide additional support at the rear of the
instrument.