3.1
SECTION 3 INITIAL CONFIGURATION
3.1 Setting passwords
The xentra has two passwords, the operator’s password gives access to CALIBRATE,
while the supervisor’s password gives access to EDIT, SETUP and CALIBRATE. The
xentra is delivered with both of the passwords set to 4000. The passwords may be
changed to any four digit number. To change the passwords follow the procedure in
Table 3.1.
Table 3.1 Changing the passwords
L MENU to obtain top level menu
CALIBRATE/SETUP
ALARMS/FAULTS
L < to move the cursor to SETUP (cursor
position shown in reverse video).
L ENTER to choose SETUP.
SET ALARM/ASSIGN
DISPLAY/UTILITY
L ? to move the cursor to DISPLAY
L < to move the cursor to UTILITY
L ENTER to choose UTILITY
ENTER PASSWORD
0000
Both user and supervisor passwords are factory
set to 4000. The cursor will be on digit furthest
to the left initially.
L > > > > to increment the first digit to 4. The
display should now be showing 4000.
L ENTER to enter the password
UTILITY 1
UTILITY 2
L ENTER to choose UTILITY 1
CLOCK/
NEW PASS
COMMS/WINDOW
L < to move the cursor to NEW PASS
L ENTER to choose NEW PASS
SUPERVISOR
OPERATOR
To change supervisor password:
L ENTER to choose SUPERVISOR.
To change operator password:
L ? to move the cursor to OPERATOR
L ENTER to choose OPERATOR.
NEW OPERATOR
(NEW SUPERVISOR)
PASS = 0000
Enter new password.
To change the value of a digit L > or ?
To change to another digit L = or <
When the value shown is correct L ENTER
NEW PASS IS ????
ACCEPT YES/NO
To accept the new password L ENTER
To reject the new password L < ENTER
SUPERVISOR
OPERATOR
To return to measurement display L MEASURE