3 ASSEMBLY AND INSTALLATION
3.3.4 Connecting the sender/receiver unit
3.3.4.1 Work to be done
Connect the connection line.
3.3.4.2 Connection of connection line
Fig. 15: Plug connector connection to onsite line
NOTE:
Only use shielded lines with twisted pairs (e.g. UNITRONIC LiYCY (TP)
4 x 2
x 0.25 mm² from LAPPKabel; not suitable for underground laying).
Cable provided by customer according to
page 29, §3.3.2
Closed
Closed
Open
Open
A
A
A -A
+24 V
Note
To open, connect the plug connector to the plug
on the sender/receiver unit.
RS485 B
RS485 A

gnd
Shield
CAN-Bus