4 START-UP AND PARAMETER SETTINGS
4.2.2.2 Addressing
For operation of the measuring system with option MCU (see “Device components”,
page 13), an address must be assigned to the sender/receiver unit.
▸ Connect the jumpers to the corresponding slots for addressing according to the
fo
llowing Table (see “Slots on processor board”, page 39).
4.2.2.3 Adapting the CAN bus baud rate
When the measuring system is operated with option SCU (see “Device components”,
page 13), it may be required to reduce the baud rate set at the factory (see “Factory set-
tings”, page 38).
The following settings are possible (slots for setting the CAN bus baud rate, see “Slots on
processor board”, page 39):
Slot Address
1 2 3 4 RS485
(for MCU connection)
CAN bus
(for SCU connection)
Jumper
connected
No No No No 1 (default) 0 → 0x10 (hex)
YesNoNoNo 1 1 → 0x11 (hex)
No Yes No No 2 2 → 0x12 (hex)
Yes Yes No No 3 3 → 0x13 (hex)
No No Yes No 4 4 → 0x14 (hex)
Yes No Yes No 5 5 → 0x15 (hex)
No Yes Yes No 6 6 → 0x16 (hex)
Yes Yes Yes No 7 7 → 0x17 (hex)
No No No Yes 8 8 → 0x18 (hex)
Yes No No Yes 1 (default) 9 → 0x19 (hex)
No Yes No Yes 1 (default) 10 → 0x1a (hex)
Yes Yes No Yes 1 (default) 11 → 0x1b (hex)
No No Yes Yes 1 (default) 12 → 0x1c (hex)
Yes No Yes Yes 1 (default) 13 → 0x1d (hex)
No Yes Yes Yes 1 (default) 14 → 0x1e (hex)
Yes Yes Yes Yes 1 (default) 15 → 0x1f (hex)
Slot 1 Slot 2 Baud rate
Jumper
connected
No No 20 kBd
Yes No 50 kBd
No Yes 100 kBd
Yes Yes 125 kBd