EasyManua.ls Logo

SICK DUSTHUNTER SB30

SICK DUSTHUNTER SB30
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
41
OPERATING INSTRUCTIONS | DUSTHUNTER SB30
8014551/1COY/3-0/
2016-08| SICK
Subject to change without notice
START-UP AND PARAMETER SETTINGS 4
4.2.3 Installing SOPAS ET
Install SOPAS ET on a laptop/PC.
Start SOPAS ET.
Following the installation instructions of SOPAS ET.
4.2.3.1 Password for SOPAS ET menus
Certain device functions are first accessible after a password has been entered.
4.2.4 Connection to the device via USB line
Recommended procedure:
1 Connect the RS485 line to the sender/receiver unit (see “DUSTHUNTER SB30 device
components”, page 14) and the laptop/PC (accessories see “Accessories”, page 69).
2 Switch the device on.
3 Start SOPAS ET.
4 “Search settings”
5 “Device family oriented search”
6 Click the desired MCU.
7 Make the settings:
Ethernet communication (always clicked)
USB communication (always clicked)
Serial communication: Click
8 Do not specify IP addresses.
9 A list of COM ports is shown.
Specify the COM port of the DUSTHUNTER.
If you do not know the COM port: see “Finding the DUSTHUNTER COM port”, page 41
10 Assign a name for this search.
11 “Finish”
4.2.4.1 Finding the DUSTHUNTER COM port
If you do not know your COM port: You can find the COM port with the Windows Device
Manager (Administrator rights are not required).
1 Disconnect the DUSTHUNTER from your laptop/PC.
2 Input: devmgmt.msc
User level Access to
0Operator
Displays measured values and system states
No password required
1 Authorized operator
Displays, inquiries as well as start-up or adjustment to customer-
specific demands and diagnosis of necessary parameters.
Preset password: sickoptic
CAN-Bus

Table of Contents

Other manuals for SICK DUSTHUNTER SB30